Vitality Life Science
WEIGHT LOSS INJECTION
reduce weight + decrease appetite
Instructions
-
Step 1:
-
1. Wash your hands with soap and water.
-
2. Pull off the pen cap.
-
Step 2:
-
1. Check that the VLS pen is full.
-
Step 3:
-
1. Take a new needle,
and tear off the paper tab.
-
2. Do not attach a new
needle to your pen until
you are ready to give your
injection.
-
Step 4:
-
1. Push the needle
straight onto the pen.
Turn until it is on
tight.
-
Step 5:
-
1. The needle is covered
by 2 caps. You must
remove both caps. If
you forget to remove both
caps, you will not inject
any medicine.
-
2. Pull off the outer
needle cap. Do not
throw it away.
-
Step 6:
-
1. Pull off the inner
needle cap and throw
it away.
-
Step 7:
-
1. Check the VLS pen
flow before the first
injection with each
new pen only.
-
2. Turn the dose selector
until the dose counter
shows the flow check symbol ▪▪▰
-
Step 8:
-
1. Hold the pen with the
needle pointing up.
Press and hold in the
dose button until the
dose counter shows 0.
The 0 must line up with
the dose pointer.
A drop of liquid will
appear at the needle tip.
-
2. If no drop appears,
repeat Step 7 above as
shown in Figure G and
Figure H up to 6 times.
-
3. If there is still no drop,
change the needle and
repeat Step 7 as shown
in Figure G and Figure H,
1 more time.
-
Do not use the pen if
a drop of liquid still
does not appear.
Contact at vls-pen@protonmail.com.
-
Step 9:
-
Select your dose
-
1. Turn the dose selector
until the dose counter
stops and shows your
desired dose.
The dashed line in the
dose counter (┇) will
guide you.
-
2. Always use the dose counter and the dose pointer to
see that your selected dose has been selected.
You will hear a “click” every time you turn the dose selector.
Do not set the dose by counting the number of clicks
you hear.
-
Step 10:
-
To see how much VLS solution is left in your pen, use the dose
counter:
-
1. Turn the dose selector
until the dose counter
stops.
-
2. If it shows 1, at
least 1 mg is left
in your pen. If the
dose counter stops
before 1 mg,
there is not enough
VLS solution left for a
full dose of 1 mg.
-
If there is not enough VLS liquid left in your pen for a
full dose, do not use it. Use a new VLS pen.
-
Step 11:
-
Inject your dose
-
1. Choose your injection
site and wipe the skin
with an alcohol swab.
Let the injection site dry
before you inject your
dose (See Figure K).
-
Step 12:
-
1. Insert the needle
into your skin as your
healthcare provider has
shown you.
-
2. Make sure you can
see the dose counter.
Do not cover it with your
fingers. This could stop
the injection.
-
Step 13:
-
1. Press and hold down
the dose button until
the dose counter
shows 0.
-
2. The 0 must line up with
the dose pointer. You may
then hear or feel a click.
-
3. Continue pressing
the dose button while
keeping the needle in
your skin.
-
Step 14:
-
1. Count 6 seconds
while keeping the
dose button pressed.
-
2. If the needle is removed
earlier, you may see a
stream of VLS solution
coming from the needle
tip. If this happens, the
full dose will not be
delivered.
-
Step 15:
-
1. Remove the needle
from your skin. You
can then release the dose
button.
-
2. If blood appears at the
injection site, press
lightly with a gauze pad
or cotton ball.
-
Do not rub
the area.
-
Step 16:
-
After your injection
-
1. Carefully remove the
needle from the pen.
Do not put the needle
caps back on the needle
to avoid needle sticks.
-
Step 17:
-
1. Place the needle in
a sharps disposal
container right away to
reduce the risk of needle
sticks.
-
Step 18:
-
1. Put the pen cap on
your pen after each use to
protect VLS pen from
light.
-
Step 19:
-
1. If you do not have a
sharps disposal container,
follow a 1-handed needle
recapping method.
Carefully slip the needle
into the outer needle cap.
Dispose of the needle in a
sharps disposal container
as soon as possible.
-
2. Never try to put the inner needle cap back on the needle. Always remove the needle from your pen.
-
3. Always dispose of the needle after each injection.